ResMed AirSense 10 Top Cover Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

What you need

  1. ResMed AirSense 10 Top Cover Replacement, Top Cover: step 1, image 1 of 2 ResMed AirSense 10 Top Cover Replacement, Top Cover: step 1, image 2 of 2
    • Remove the water tub by gently pulling it out.

  2. ResMed AirSense 10 Top Cover Replacement: step 2, image 1 of 3 ResMed AirSense 10 Top Cover Replacement: step 2, image 2 of 3 ResMed AirSense 10 Top Cover Replacement: step 2, image 3 of 3
    • Pry off the face plate by using your finger or an opening tool.

    • Use a moderate amount of force when prying open.

    • Remove the face plate from the top cover.

    • There will be a clicking sound on either side of the divot when assembling and disassembling.

  3. ResMed AirSense 10 Top Cover Replacement: step 3, image 1 of 2 ResMed AirSense 10 Top Cover Replacement: step 3, image 2 of 2
    • Unscrew and remove the two 11.8 mm TR10 Torx screws that were revealed from the face plate being removed.

  4. ResMed AirSense 10 Top Cover Replacement: step 4, image 1 of 2 ResMed AirSense 10 Top Cover Replacement: step 4, image 2 of 2
    • Unscrew and remove the two 11.8 mm TR10 Torx screws from the bottom of the device.

  5. ResMed AirSense 10 Top Cover Replacement: step 5, image 1 of 3 ResMed AirSense 10 Top Cover Replacement: step 5, image 2 of 3 ResMed AirSense 10 Top Cover Replacement: step 5, image 3 of 3
    • Remove the top cover from the rest of the device.

Conclusion

To reassemble your device, follow these instructions in reverse order.