What you need

  1. Philips Respironics V60 Top Cover Replacement, shut down the ventilator: step 1, image 1 of 2 Philips Respironics V60 Top Cover Replacement, shut down the ventilator: step 1, image 2 of 2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    Ask FixBot
    • Disconnect the ventilator from the AC power.

    Ask FixBot
    • Unplug the ventilator

    Ask FixBot
  2. Philips Respironics V60 Top Cover Replacement: step 4, image 1 of 1
    • Disconnect the internal battery from the power harness connector.

    Ask FixBot
  3. Philips Respironics V60 Top Cover Replacement, Top Cover: step 5, image 1 of 1
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

    Ask FixBot
  4. Philips Respironics V60 Top Cover Replacement: step 6, image 1 of 1
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

    Ask FixBot
Conclusion

To reassemble your device, follow these instructions in reverse order.

Rie Kuraoka

Member since: 01/23/20

913 Reputation

7 Guides authored

Team

Cal Poly, Team S7-G10, Paton Spring 2020 Member of Cal Poly, Team S7-G10, Paton Spring 2020

CPSU-PATON-S20S7G10

2 Members

9 Guides authored

0 Guide Comments

Add Comment

View Statistics:

Past 24 Hours: 1

Past 7 Days: 14

Past 30 Days: 62

All Time: 546