What you need

  1. Ohmeda Engström Compressor Replacement, Compressor: step 1, image 1 of 1
    • Ensure the device is powered off and unplugged

    • Disconnect any wires or components that are plugged in prior to repair

  2. Ohmeda Engström Compressor Replacement: step 2, image 1 of 1
    • Tap the back end of a flat head screwdriver with a hammer to wedge the side trim pieces away from the machine.

  3. Ohmeda Engström Compressor Replacement: step 3, image 1 of 1
    • Use the flat head screwdriver to pry off and remove the side panels

    • When prying off the panels, do so carefully so that you do not to scratch or damage them

    • When reassembling the side panels, make sure that the connectors snap back into place.

    • Remove the mounting hardware from the compressor on the cart

  4. Ohmeda Engström Compressor Replacement: step 4, image 1 of 1
    • Screw in the service tool to assist in lifting the compressor and sliding out of the cart

Conclusion

To reassemble your device, follow these instructions in reverse order.