Skip to main content

Free Shipping on Domestic Orders $75+

How to power down a Hamilton C2

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.
    1. How to power down a Hamilton C2, shut down the ventilator: step 1, image 1 of 2 How to power down a Hamilton C2, shut down the ventilator: step 1, image 2 of 2
      • Press and quickly release the power switch to open standby window in the screen.

      • Activate the standby mode by pressing the Active standby button in the screen.

      • Press and hold the switch for greater than 3 seconds to shut down the ventilator.

      • If there is a technical fault, press and hold the switch for greater than 10 seconds.

      Ask FixBot

      Add Comment

      • Disconnect the ventilator from the AC or DC power.

      Ask FixBot

      Add Comment

Finish Line

Cancel: I did not complete this guide.