Time Required: 10 minutes
Difficulty: Easy
Hamilton-T1 Oxygen Cells Replacement
Medical Equipment Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

Introduction

Go to step 1 ↓

This guide is created for the replacement of oxygen cells in the Hamilton-T1 Ventilator. Replacing the oxygen cells is a crucial step in maintaining and keeping a working ventilator.

NOTE:

  • Replace the oxygen cell with genuine Hamilton Medical parts only; otherwise, oxygen measurement will not function. To prevent a permanent alarm make sure to use special Hamilton Medical oxygen cells only.
  • To prevent leakage within the ventilator, make sure an oxygen cell is installed at all times, even if you use an external monitor or disable oxygen monitoring.

What you need

Tools

    1. Hamilton-T1 Oxygen Cells Replacement, Oxygen Cells: step 1, image 1 of 1
      • Flip device around until the back panel is exposed.

      Ask FixBot

      Add Comment

    2. Hamilton-T1 Oxygen Cells Replacement: step 2, image 1 of 1
      • Remove the screw in the bottom center of the panel, and then take the panel off.

      Ask FixBot

      Add Comment

    3. Hamilton-T1 Oxygen Cells Replacement: step 3, image 1 of 1
      • Disconnect the oxygen cell by removing the tube.

      • Then twist the cell counter-clockwise and pull to remove the oxygen cell.

      • Install the new oxygen cell and put the back cover on.

      Ask FixBot

      Add Comment

      • After the cell is installed. Run the oxygen cell calibration.

      • Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.

      Ask FixBot

      Add Comment

Conclusion

After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.

Cancel: I did not complete this guide.

Attached Documents

Tyler

Member since: 04/09/20

114 Reputation

1 Guide authored

Team

Cal Poly, Team S11-G24, Paton Spring 2020 Member of Cal Poly, Team S11-G24, Paton Spring 2020

CPSU-PATON-S20S11G24

3 Members

36 Guides authored

0 Guide Comments

Add Comment

View Statistics:

Past 24 Hours: 2

Past 7 Days: 16

Past 30 Days: 86

All Time: 4,428