Fisher Paykel Neopuff Valve System Replacement

Medical Equipment
Neither iFixit nor this repair guide is endorsed by or affiliated with the medical device manufacturer.

What you need

  1. Fisher Paykel Neopuff Valve System Replacement, Valve System: step 1, image 1 of 1
    • Remove the back cover, fixed by four screws.

    Ask FixBot

    Add Comment

    • Disconnect the tube from the manometer.

    Ask FixBot

    Add Comment

    • Remove the manometer by unscrewing the two retaining nuts.

    Ask FixBot

    Add Comment

    • Fit the existing manometer into the new valve assembly panel.

    Ask FixBot

    Add Comment

    • Tighten the retaining nuts.

    Ask FixBot

    Add Comment

    • Reconnect the manometer tube.

    Ask FixBot

    Add Comment

    • Refit the front panel to the back cover with the four screws from step 1.

    Ask FixBot

    Add Comment

  2. Fisher Paykel Neopuff Valve System Replacement: step 8, image 1 of 1
    • Carry out the valve system performance test described above. It is recommended to record the lot number of the new valve assembly onto the Set-up and Maintenance Checklist.

    Ask FixBot

    Add Comment

Conclusion

To reassemble your device, follow these instructions in reverse order.