Background

The GLUCOCARD 01 Blood Monitoring System is a monitor launched by Arkray and approved by the FDA in 2008. There is a companion model, the Glucocard 01 Mini, which was released in the following year.

The blood monitoring system was designed to measure blood glucose levels in diabetic persons in their homes to monitor the effectiveness of their treatment. It is designed to be used by a single person per device, and saves up to 360 dated results. It measures whole blood taken from the fingertips or palm of the user.

The entire system comes with the meter, its carrying case, a user manual and guide, a logbook for results, a lancing device, control solution, and test strips. The meter also requires a battery to be inserted on the back side, which typically grants approximately 2,000 uses of the meter before the battery is replaced.

While Arkray has released multiple different models of blood glucose monitoring systems, the 01 remains in use due to its simplicity and quickness, as it usually delivers results within 7 seconds, which are timed and dated within the device itself. It is also an autocoded meter, meaning the user does not have to manually calibrate the device, allowing for less issues to be present in the results from human error.

Identification

The Glucocard 01 Blood Monitoring System is a compact meter with a gray rectangular body. It features a display screen for glucose level results, with a test strip port on the bottom of the device and arrow buttons to the right side of the device. To positively identify the device, in the space above the display screen there is text that reads “Glucocard 01” in capital letters, and the Arkray label just below the display screen.

Technical Specifications

Test Strips: GLUCOCARD 01 SENSOR PLUS.

Result Range: 20-600 mg/dL.

Calibration: Plasma referenced.

Sample Size: Minimum 0.3 μL.

Blood Source: Fresh capillary whole blood.

Hematocrit Range: 30% - 55%.

Test Time: 7 seconds.

Assay Method: Biosensor, Glucose Oxidase (Aspergillus niger sourced).

Power Source: One 3-volt lithium battery (CR2032 or DL2032).

Battery Life: Approximately 2,000 tests.

Screen Type: Liquid Crystal Display (LCD).

Memory: 360 results with time & date stamp.

Average: 14-day or 30-day averaging.

Automatic Shut Off: 3 – 5 minutes after last user action.

Size: 2" x 3.9" x 0.5".

Weight: 1.6 oz (with battery).

Operating Ranges: 50°F-104°F (10°C-40°C), 20% - 80% relative humidity.

Altitude: 10,000 ft. (3,048 meters).

Unit of Measurement: mg/dL.

Additional Information

Glucocard BLOOD GLUCOSE MONITORING SYSTEM

Arkray Product Information

Diabetic Plaza Glucocard Meter Features

FDA Glucocard-01 Review Memorandum

Author

with

eCuration

Member since: 05/19/17

478,744 Reputation

396 Guides authored

Team

iFixit Member of iFixit

Staff

119 Members

72,966 Guides authored

View Statistics:

Past 24 Hours: 0

Past 7 Days: 4

Past 30 Days: 32

All Time: 293